Yes, the probes, the wire and the metal clip can be cleaned and disinfected in a washer disinfector.

The procedure below has been validated according to EN ISO 15883:

  • Use enzymatic or low alkaline, suited for medical devices, non-corrosive, neutral cleaning agent with minimal foaming characteristics, suited for medical devices.
  • Use drinking water quality for cleaning and deionized or purified water for final rinse/disinfection.
  • Load the instruments in an open position in the washer-disinfector. Use a suitable instrument tray.
  • Pre-wash in cold water, 2 x 2 min
  • Main wash with detergent at minimum 55°C, 10 min
  • Rinse in warm water, 2x 1 min.
  • Final rinse/disinfection in distilled or de-mineralized water at 90°C, minimum 1 min.
  • Drying at 110° C, minimum 15 min.